Related Experiment Video
Updated: Dec 23, 2025

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
A large buoy-based radioactivity monitoring system for gamma-ray emitters in surface seawater
Jong-In Byun1, Seok-Won Choi1, Myeong-Han Song1
1Korea Institute of Nuclear and Safety, 62 Gwahak-ro Yuseong-gu, 34142, Daejeon, Republic of Korea.
Abstract:
A buoy (shallow water light type) -based in situ gamma-ray spectrometry system with a 7.6 cmØ × 7.6 cm NaI(Tl) detector for remote real-time monitoring of gamma-ray emitting radionuclides in surface seawater is presented. To convert measured count rates to radioactivity, the full energy peak efficiency of the detector for radionuclides in seawater was estimated using Monte Carlo simulation with the MCNP code. The efficiency calibration was validated by comparing the results with a sampling analysis of 40K in seawater at the sites where the monitoring systems were deployed. The minimum detectable activity of the system for 137Cs, 134Cs and 131I with gamma-ray measurement time is discussed.
Related Concept Videos
Buoyancy and Stability for Submerged and Floating Bodies
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET