Comparing image quality of single- and dual-energy computed tomography of the brain
Doris Dodig1, Slavica Kovačić1, Zrinka Matana Kaštelan1
1Radiology Department, Clinical Hospital Center Rijeka, Croatia.
Purpose:
Weighted average dual-energy computed tomography (DE-CT) reconstructions are considered a proxy of standard CT images of the brain, recommended for routine clinical use and used as a reference standard in DE-CT research. However, their image quality has not been assessed, which was the aim of our study.
Methods:
Images from 81 consecutive patients who underwent both non-contrast single-energy (SE)-CT and DE-CT of the brain on the same scanner were retrospectively evaluated. Attenuation values (HU) and SD of grey matter/white matter (GM/WM) pairs, along with SD in the posterior fossa and subcalvarial region were measured. Four readers evaluated image noise, GM/WM contrast, posterior fossa and subcalvarial artefacts, as well as overall image quality.
Results:
Weighted average DE-CT GM and WM HU were significantly lower and noise higher compared to SE-CT (GM HU 36.46 v. 41.82; WM HU 28.18 v. 29.94; GM SD 2.93 v. 2.49; and WM SD 3.16 v. 2.44, all p < 0.0001). After correcting the measured SE-CT noise for 37% higher acquisition dose, DE-CT GM noise became significantly lower (2.93 v. 3.11, p = 0.0121). Measured and dose corrected SE-CT GM/WM contrast-to-noise ratio was superior to weighted average DE-CT (3.42 and 2.74 v. 1.95, both p < 0.0001). Weighted average DE-CT had significantly less artifacts on qualitative analysis.
Conclusion:
Weighted average DE-CT images of the brain yield less artefacts at 37% dose reduction and lower noise at SE-CT equivalent dose. Dose-adjusted GM/WM contrast-to-noise ratio of weighted average DE-CT with 0.4 weighting factor remains inferior to SE-CT images.
More Related Videos
07:28Studying Metabolic Brain Connectivity Using 2-Deoxy-2-[18F]Fluoro-D-Glucose Dynamic Positron Emission Tomography at the Single-subject Level
Published on: January 24, 2025
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...