To measure T1 of short T2 species using an inversion recovery prepared three-dimensional ultrashort echo time (3D
Zhao Wei1, Ya-Jun Ma2, Hyungseok Jang2
1Department of Radiology, University of California San Diego, CA, United States; University of Chinese Academy of Sciences, Beijing, China; Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, China.
Purpose:
To demonstrate the feasibility of a new method for measuring T1 of short T2 species based on an adiabatic inversion recovery-prepared three-dimensional ultrashort echo time Cones (3D IR-UTE-Cones) sequence.
Methods:
T1 values for short T2 species were quantified using 3D IR-UTE-Cones data acquired with different repetition times (TRs) and inversion times (TIs). An inversion efficiency factor Q was introduced into the fitting model to accurately calculate T1 values for short T2 species. Experiments were performed on twelve MnCl2 aqueous solution phantoms with a wide range of T1 values and T2* values on a 3 T clinical MR system to verify the efficacy of the proposed method. For comparison, a variable flip angle UTE (VFA-UTE) sequence, a variable TR UTE (VTR-UTE) sequence, and a conventional 2D IR fast spin echo (IR-FSE) sequence were also used to quantify T1 values of those phantoms. T1 values were compared between all performed sequences.
Results:
The proposed 3D IR-UTE-Cones method provided higher contrast images of short T2 phantoms and measured much shorter T1 values than the VFA-UTE, VTR-UTE and 2D IR-FSE methods. T1 values as short as 2.95 ms could be measured by the 3D IR-UTE-Cones sequence. The 3D IR-UTE-Cones methods with different TRs were applied to different ranges of T1 measurement, and the scan time was significantly decreased by using 5 TIs along the recovery curves to perform fitting with comparable accuracy.
Conclusion:
The 3D IR-UTE-Cones sequence could accurately measure short T1 values while providing high contrast images of short T2 species.