Low-Dose Image-Guided Pediatric CNS Radiation Therapy: Final Analysis From a Prospective Low-Dose Cone-Beam CT
Sara R Alcorn1, Xian Chiong Zhou2, Casey Bojechko3
1The Johns Hopkins Hospital, Baltimore, MD, USA.
A new lower-dose cone-beam computed tomography protocol effectively localizes targets for pediatric brain tumor radiotherapy, minimizing radiation exposure. This method proved accurate across diverse patient groups, with images rated adequate for localization in over 99% of cases.
Area of Science:
- Radiotherapy
- Medical Imaging
- Pediatric Oncology
Background:
- Image-guided radiotherapy (IGRT) aims to improve treatment accuracy.
- Lower-dose imaging protocols are crucial for minimizing radiation exposure in pediatric patients.
- Cone-beam computed tomography (CBCT) is a key technology in IGRT.
Purpose of the Study:
- To evaluate a lower-dose CBCT protocol for central nervous system (CNS) IGRT in pediatric patients.
- To assess the setup accuracy and radiation dose of the lower-dose CBCT protocol.
- To determine the suitability of the protocol across a diverse pediatric population.
Main Methods:
- A prospective, multinational study involving seven institutions.
- A lower-dose CBCT protocol (0.7 mGy average dose) was used for patients ≤21 years.
- Setup accuracy was assessed by comparing table shifts from surface lasers versus CBCT, calculating vector magnitudes and estimating errors (group mean, interpatient, interinstitution, random).
Main Results:
- 96 patients with 2179 CBCT acquisitions were analyzed (median age 9 years).
- Setup accuracy vector magnitudes were 3.13 mm (group mean), 3.02 mm (interpatient), 1.64 mm (interinstitution), and 1.48 mm (random error).
- No significant clinical factors (age, gender, performance status, etc.) impacted setup accuracy; >99% of images were adequate for localization.
Conclusions:
- The lower-dose CBCT protocol achieved setup accuracy comparable to clinical target volume expansions in CNS radiotherapy.
- No significant clinical predictors of setup accuracy were identified, supporting broad applicability.
- This lower-dose CBCT protocol is suitable for diverse pediatric brain tumor patients, balancing accuracy and reduced radiation exposure.