Related Experiment Video
Updated: Dec 23, 2025

Design of a Biocompatible Drug-Eluting Tracheal Stent in Mice with Laryngotracheal Stenosis
Published on: January 21, 2020
Preclinical evaluation of a pediatric airway stent for tracheobronchomalacia
Abhijit Mondal1, Junhyoung Ha2, Vickie Y Jo3
1Department of Cardiovascular Surgery, Boston Children's Hospital, Boston, Mass.
Objectives:
We sought to demonstrate in an animal model that helical stents made from a nickel titanium alloy called nitinol (NiTi) and designed for malacic airways could be delivered and removed without significant trauma while minimally impeding mucus clearance during the period of implantation.
Methods:
Stents were delivered and removed from the tracheas of healthy 20 kg swine (n = 5) using tools designed to minimize trauma. In 4-week experiments, the stents were implanted on day 0, removed after 3 weeks, and swine were put to death after 4 weeks. Weekly bronchoscopies, radiographs, and mucus clearance examinations were performed in vivo. Hematoxylin and eosin staining and scanning electron microscopy imaging were used to evaluate foreign body response, tracheal tissue reaction, and damage and to measure unciliated regions.
Results:
In all in vivo experiments, the stent was implanted and removed atraumatically. Mucus clearance was maintained throughout the experiment period. Hematoxylin and eosin-stained slides showed that foreign body response and tracheal tissue damage were localized to the stented subsections. Tracheal tissue reaction and damage was further restricted to the epithelium and submucosal layers. Scanning electron microscopy imaging revealed that the cilia were absent only over the contact area between the trachea and the wire forming the helical stent.
Conclusions:
Helical nitinol stents designed to provide radial support for malacic airways were well tolerated in a porcine model, providing for mucus clearance while also enabling atraumatic removal.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...