Related Experiment Video
Updated: Dec 23, 2025

Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
Initial experience in self-monitoring of intraocular pressure
Emily McGarva1, Jane Farr1, Priya Dabasia2
1Moorfields Eye Hospital NHS Foundation Trust, London, UK.
Background/Aims:
Diurnal variation in intraocular pressure (IOP) is a routine assessment in glaucoma management. Providing patients the opportunity to perform self-tonometry might empower them and free hospital resource. We previously demonstrated that 74% of patients can use the Icare® HOME tonometer. This study further explores Icare® HOME patient self-monitoring.
Methods:
Patients were trained by standard protocol to use the Icare® HOME rebound tonometer. Patient self-tonometry was compared to Goldmann applanation tonometry (GAT) over one clinical day. Following this, each patient was instructed to undertake further data collection that evening and over the subsequent two days.
Results:
Eighteen patients (35 eyes) participated. Good agreement was demonstrated between GAT and Icare® HOME for IOPs up to 15 mm Hg. Above this IOP the Icare® tended to over-read, largely explained by 2 patients with corneal thickness >600 um. The mean peak IOP during 'clinic hours' phasing was 16.7 mm Hg and 18.5 mm Hg (p = 0.24) over three days. An average range of 5.0, 7.0 and 9.8 mm Hg was shown during single day clinic, single day home and three day home phasing respectively (p =<0.001). The range of IOP was lower in eyes with prior trabeculectomy (6.1 mm Hg vs 12.2 mm Hg). All patients undertook one reading in the early morning at home with an average of 4.8 readings during, and 3.1 readings after office hours.
Conclusions:
This small study shows that self-tonometry is feasible. The findings from home phasing demonstrated higher peak and trough IOPs, providing additional clinical information. Home phasing is a viable alternative. The cost-effectiveness of this approach has yet to be addressed.