Related Experiment Video
Updated: Aug 12, 2026

Quantifying X-Ray Fluorescence Data Using MAPS
Published on: February 17, 2018
ED-XRF: a promising method for accurate and rapid quantification of metals in a bacterial matrix
Céline Lelièvre1,2, Asmaa Rouwane3,2, Isabelle Poirier3,2
1UFR des Sciences Université de Caen Normandie, Cherbourg-en-Cotentin, France.
Abstract:
ABSTRACTThe remediation of metal-polluted water using bacterial biofilms is a promising technology. In order to help its development, the present study aims to evaluate the feasibility to utilize XRF spectrometry for accurate and rapid measurement of metal concentrations in bacterial biofilms used in treatment plants. For that purpose, an ED-XRF spectrometer was used to measure Cd, Cu, Fe, Mn, Ni and Zn concentrations within a matrix of marine bacteria Pseudomonas fluorescens BA3SM1 and its metabolites. Contaminated and control cultures of the strain BA3SM1 were dried and crushed, then analysed by ED-XRF. The LOD value of the analysed metals was between 2.08 and 10.5 µg g-1. Metal concentrations were also measured by ICP-AES or ICP-MS to support ED-XRF results. The two techniques showed a good linear correlation with a slope of at least 0.949 and R2 of at least 0.985. These results confirm the possibility to measure metal contents by ED-XRF in bacterial matrices.