New retinal tack designs: an analysis of retention forces in human scleral tissue
Markus Schulze Schwering1, Theo Oltrup2, Kai Sinan Rückheim3
1Centre for Ophthalmology, University Eye Hospital, Tübingen, Germany.
Purpose:
The study aimed to construct a new retinal tack design with high retention forces to prevent spontaneous disentanglement in cases of complicated retinal surgery.
Methods:
Six new forms for the peak of a retinal tack were developed using computer-aided design (CAD); then a prototype was produced for each model. Finally, standardised design testing was conducted using human (ex vivo) sclera by logging 15 consecutive measurements for each model.
Results:
Seven different models underwent pull-out testing (six new models and the original tack model), but two tack models (Model 4, Model 5) failed to penetrate the human tissue. The highest pull-out forces (median) were measured for Model 3, followed by Model 6, Model 2 and Model 1. The original Heimann tack (Model H) was found to have the lowest retention forces.
Conclusion:
The different tack designs altered the penetration and holding forces. The retention forces of the proposed peak design led to a significant increase in the retention forces that were more than twice as high as those in the original Heimann Model.