MRI-LINAC beam profile measurements using a plastic scintillation dosimeter
Levi Madden1, James Archer2, Enbang Li2
1Centre for Medical Radiation Physics, University of Wollongong, NSW 2522, Australia; Ingham Institute for Applied Medical Research, Liverpool, NSW 2170, Australia.
Summary
Plastic scintillation dosimeters (PSDs) show promise for MRI-LINAC dosimetry, achieving high agreement with film measurements across various criteria. This research validates PSDs for accurate dose assessment in advanced radiotherapy.
Area of Science:
- Medical Physics
- Radiotherapy Physics
- Radiation Dosimetry
Background:
- Plastic scintillation dosimeters (PSDs) offer advantageous properties for linear accelerator (LINAC) dosimetry.
- Their potential for magnetic resonance imaging-LINAC (MRI-LINAC) dosimetry requires further investigation.
- Limited research exists on PSD performance specifically within MRI-LINAC environments.
Purpose of the Study:
- To evaluate the dosimetric performance of an in-house plastic scintillation dosimeter (PSD) with an in-line MRI-LINAC.
- To compare PSD beam profiles against film measurements.
- To assess the accuracy and reliability of PSDs for relative dosimetry in MRI-LINAC applications.
Main Methods:
- An in-house PSD was utilized to acquire 8 beam profiles from an in-line MRI-LINAC.
- Film dosimetry was employed as a reference for comparison.
- One-dimensional global gamma indices (γ) and pass rates were calculated using 1%/1 mm, 2%/2 mm, and 3%/3 mm criteria.
Main Results:
- Mean global gamma pass rates reached 85.8% (1%/1 mm), 97.5% (2%/2 mm), and 99.4% (3%/3 mm).
- Gamma failures were predominantly observed in penumbral regions, with PSD penumbra widths showing agreement with film within measurement uncertainties.
- Dose differences between PSD and film remained within 2.2% globally in central regions and 1.5% in out-of-field regions.
Conclusions:
- The investigated PSD demonstrates reliable performance for relative dosimetry in MRI-LINAC setups.
- Achieved agreement levels are comparable to other established dosimeters used in MRI-LINAC relative dosimetry.
- PSDs are a viable option for dosimetry in advanced MRI-LINAC radiotherapy systems.