Determinants of the difficulty of leadless pacemaker implantation
Christophe Garweg1,2, Bert Vandenberk1,2, Stefaan Foulon2
1Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.
Background:
The Micra Transcatheter Pacing System is implanted directly in the right ventricle (RV) through the femoral vein using a steerable transcatheter delivery system. The present study was done to identify determinants of difficult leadless pacemaker implant procedures including operator, patient, and RV anatomical characteristics.
Methods:
All patients who underwent a Micra implant from July 2015 to December 2018 at our center were analyzed. From an RV angiogram acquired during implantation, RV geometry including systolic and diastolic volumes and ejection fraction was characterized. The presence of septomarginal trabeculation was noted.
Results:
One hundred twenty-six patients (mean age: 79 ± 11 years old, mostly male: 77%) were enrolled. Mean Micra RV implant procedure time was 24 ± 23 min, with 1.7 ± 1.3 deployments of the device. No significant change in implant procedure time was observed after the first 30 implants. Eleven patients had a prominent septal component of the septomarginal trabeculation in the RV. Univariate analysis showed that the procedure time was positively correlated with the presence of a prominent septal component of the septomarginal trabeculation (P < .001) or an episode of heart failure (P = .02) and negatively correlated with the number of procedures performed by the operator (P < .001). After multivariable analysis, only the presence of a prominent septal component of the septomarginal trabeculation (P < .001) and the number of procedures performed by the operator (P < .001) were associated with the implant procedure time.
Conclusions:
In our experience, implant procedure time of a Micra leadless pacemaker depended on the presence of a prominent septal component of the septomarginal trabeculation and operator experience.