Related Experiment Video
Updated: Dec 22, 2025

Author Spotlight: 3D Movement Assessment of Maxillary Posterior Teeth in Clear Aligner Treatment
Published on: February 23, 2024
Comparison of two computerized occlusal analysis systems for indicating occlusal contacts
Min-Young Jeong1, Young-Jun Lim2, Myung-Joo Kim2
1Department of Prosthodontics, School of Dentistry, Seoul National University, Seoul, Republic of Korea.
Accura demonstrates comparable performance to the T-scan in identifying occlusal contacts. This computerized occlusal analysis system shows potential as an alternative for dental diagnostics.
Area of Science:
- Dental Technology
- Biomedical Engineering
- Occlusion Analysis
Background:
- Accurate identification of occlusal contacts is crucial for dental diagnostics and treatment planning.
- Existing technologies like T-scan provide valuable occlusal data.
- The development of new systems aims to enhance diagnostic capabilities and patient outcomes.
Purpose of the Study:
- To evaluate and compare the performance of the Accura system against the T-scan system.
- To determine Accura's efficacy in indicating occlusal contacts during maximum intercuspation.
Main Methods:
- Twenty-four subjects' maxillary dental casts were scanned and digitally processed.
- Occlusal contacts were recorded using both T-scan and Accura during masticatory cycles.
- Data from both systems were overlapped onto the cast models for comparative analysis.
- Statistical analysis, including McNemar's test, was employed to assess differences.
Main Results:
- No statistically significant differences were observed between the T-scan and Accura methods (P > .05).
- Accura exhibited an accuracy of 75.8%, sensitivity of 82.1%, and specificity of 60.1%.
- Positive predictive value (PPV) was 82.9%, and negative predictive value (NPV) was 60.1%.
Conclusions:
- Accura presents a viable alternative computerized occlusal analysis system.
- The system effectively indicates occlusal contacts at maximum intercuspation.
- Further research may explore Accura's clinical applications and long-term reliability.
More Related Videos
08:47Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024