Related Experiment Video
Updated: Dec 22, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill
Jong Won Kim1, Sang O Park1, Kyeong Ryong Lee1
1Department of Emergency Medicine, Konkuk University Medical Center, Konkuk University School of Medicine, Seoul, Republic of Korea.
Objective:
The aim of this study was to test whether Amflow® (a newly designed portable ventilation feedback device) can assist rescuers in delivering target tidal volume (V T) and respiration rate (RR) during self-inflating bag (SB) ventilations in various clinical scenarios.
Method:
This was a simulation study with a prospective cross-over design. A total of 40 trained participants who underwent training for SB ventilation were recruited. Using a SB with or without Amflow® alternately, participants delivered ventilations to test lungs connected to a gas flow analyser in each of three different scenarios: acute respiratory distress syndrome (ARDS; 315-385 ml ranges for 350 ml target V T, with 20 breaths/min); cardiopulmonary resuscitation (CPR; 450-550 ml ranges for 500 ml target V T with 10 breaths/min); and adult head trauma (630-770 ml ranges for 700 ml target V T with 15 breaths/min).
Results:
The feedback group (SB with Amflow®) demonstrated a significantly higher percentage of delivering the appropriate V T ranges than the no-feedback group for both ARDS (58.6% versus 23.5%, respectively) and CPR (85.4% versus 41.0%, respectively) (all p < 0.05). However, there was no significant difference between the two groups in the percentage of delivering the appropriate V T ranges in head trauma patients (65.9% versus 68.3%, respectively; p=0.092). In all three scenarios, a higher percentage of target RR delivered was achieved in the feedback group (88.3%, 99.2%, and 96.3%, respectively) compared with the no-feedback group (5.8%, 12.5%, and 10.0%, respectively) (all p < 0.05).
Conclusion:
The Amflow® device could be useful for rescuers in delivering SB ventilation with appropriate V T and RR simultaneously in various critical situations, except for clinical cases that demand greater delivered V T.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: