Retrospectively ECG-gated helical vs. non-ECG-synchronized high-pitch CTA of the aortic root for TAVI planning

Barbora Horehledova1,2, Casper Mihl1,2, Ellen Boswijk1,2

  • 1Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.

Plos One
|May 13, 2020
PubMed

Insights

Optimal imaging for transcatheter aortic valve implantation (TAVI) is crucial. Systolic imaging is necessary for accurate aortic annulus measurements, impacting prosthesis size selection in nearly half of TAVI candidates.

Area of Science:

  • Cardiovascular imaging
  • Interventional cardiology
  • Radiology

Background:

  • Multidetector computed tomography (MDCT) is vital for pre-transcatheter aortic valve implantation (TAVI) assessment.
  • Optimal pre-TAVI imaging protocols lack consensus, potentially causing discrepancies in aortic annulus measurements and prosthesis selection.

Purpose of the Study:

  • To evaluate differences in aortic annulus measurements using systolic, diastolic, and non-ECG-synchronized imaging.
  • To assess the impact of imaging methods on TAVI prosthesis size selection.

Main Methods:

  • Retrospective analysis of 50 TAVI candidates using ECG-gated CT angiography and non-ECG-synchronized CT.
  • Aortic root dimensions assessed across the cardiac cycle (0-100% R-R interval) and on non-ECG-synchronized scans.
  • Comparison of measurements using ANOVA and Bland-Altman analysis.

Main Results:

  • Max-systolic measurements were significantly larger than end-diastolic and non-ECG-synchronized measurements (p<0.001).
  • Prosthesis size selection varied across methods in 48-62% of patients.
  • Significant and clinically relevant discrepancies were observed between imaging strategies.

Conclusions:

  • Systematic differences in aortic annular dimensions based on imaging timing are clinically significant.
  • Imaging strategy impacts prosthesis selection in a substantial portion of TAVI candidates.
  • Systolic imaging is essential for accurate assessment of maximal annular dimensions and minimum annulus-coronary ostia distances for optimal TAVI prosthesis selection.
Abstract