Spatial resolution compensation by adjusting the reconstruction kernels for iterative reconstruction images of
Koichi Sugisawa1, Katsuhiro Ichikawa2, Atsushi Urikura3
1Department of Radiological Technology, Keio University Hospital, 35 Shinanomachi, Shinjuku, Tokyo 160 8582, Japan; Division of Health Sciences, Graduate School of Medical Science, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa 920 0942, Japan.
Purpose:
Hybrid iterative reconstruction (IR) is useful to reduce noise in computed tomography (CT) images. However, it often decreases the spatial resolution. The ability of high spatial resolution kernels (harder kernels) to compensate for the decrease in the spatial resolution of hybrid IRs was investigated.
Methods:
An elliptic cylindrical phantom simulating an adult abdomen was used. Two types of rod-shaped objects with ~330 and ~130 HU were inserted to simulate contrasts of arteries in CT angiography. Two multi-slice CT systems were used to scan the phantoms with 120 kVp and scan doses of 20 and 10 mGy. The task transfer functions (TTFs) were measured from the circular edges of the rod images. The noise power spectrum (NPS) was measured from the images of the water-only section. The CT images were reconstructed using a filtered back projection (FBP) with baseline kernels and two levels of hybrid IRs with harder kernels. The profiles of the clinical images across the aortic dissection flaps were measured to evaluate actual spatial resolutions.
Results:
The TTF degradation of each hybrid IR was recovered by the harder kernels, whereas the noise reduction effect was retained, for both the 20 and 10 mGy. The profiles of the dissection flaps for the FBP were maintained by using the harder kernels. Even with the best combination of hybrid IR and harder kernel, the noise level at 10 mGy was not reduced to the level of FBP at 20 mGy, suggesting no capability of a 50% dose reduction while maintaining noise.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...