Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Contact feedback improves 1-year outcomes of remote magnetic navigation-guided ischemic ventricular tachycardia
Anna Maria Elisabeth Noten1, Astrid Armanda Hendriks1, Sing-Chien Yap2
1Department of Cardiology, Erasmus MC, University Medical Center, Rotterdam, the Netherlands; Department of Cardiology, Onze Lieve Vrouwe Gasthuis, Amsterdam, the Netherlands.
Introduction:
Remote magnetic navigation (RMN)-guided catheter ablation (CA) is a feasible treatment option for patients presenting with ischemic ventricular tachycardia (VT). Catheter-tissue contact feedback, enhances lesion formation and may consequently improve CA outcomes. Until recently, contact feedback was unavailable for RMN-guided CA. The novel e-Contact Module (ECM) was developed to continuously monitor and ensure catheter-tissue contact during RMN-guided CA.
Objective:
The present study aims to evaluate the effect of ECM implementation on acute and long-term outcomes in RMN-guided ischemic VT ablation.
Method:
This retrospective, two-center study included consecutive ischemic VT patients undergoing RMN-guided CA from 2010 to 2017. Baseline clinical data, procedural data, including radiation times, and acute success rates were compared between CA procedures performed with ECM (ECM+) and without ECM (ECM-). One-year VT-free survival was analyzed using Cox-proportional hazards models, adjusting for potential confounders: age, left ventricular function, VT inducibility at baseline and substrate based ablation strategy.
Results:
The current study included 145 patients (ECM+ N = 25, ECM- N = 120). Significantly lower fluoroscopy times were observed in the ECM+ group (9.5 (IQR 5.3-13.5) versus 12.5 min (IQR 8.0-18.0), P = 0.025). Non-inducibility of the clinical VT at the end of procedure was observed in 92% ECM+ versus 72% ECM- patients (P = 0.19). ECM guidance was associated with significantly lower VT-recurrence rates during 1-year follow-up (16% ECM+ versus 40% ECM-; multivariable HR 0.29, 95%-CI 0.10-0.69, P = 0.021, reference group: ECM-).
Conclusion:
Contact feedback by the ECM further decreases fluoroscopy exposure and improves VT-free survival in RMN-guided ischemic VT ablation.