Impact of dead time on quantitative 177Lu-SPECT (QSPECT) and kidney dosimetry during PRRT
Alessandro Desy1,2, Guillaume F Bouvet1,2, Andrea Frezza3
1Cancer Research Centre and Department of Radiology and Nuclear Medicine, Université Laval, Quebec City, Canada.
Background:
Dead time may affect the accuracy of quantitative SPECT (QPSECT), and thus of dosimetry. The aim of this study was to quantify the effect of dead time on 177Lu-QSPECT and renal dosimetry following peptide receptor radionuclide therapy (PRRT) of neuroendocrine tumours.
Methods:
QSPECT/CT was performed on days 1 and 3 during 564 personalized 177Lu-octreotate cycles in 166 patients. The dead-time data for each scanning time point was compiled. The impact of not correcting QSPECT for the dead time was assessed for the kidney dosimetry. This was also estimated for empiric PRRT by simulating in our cohort a regime of 7.4 GBq/cycle.
Results:
The probability to observe a larger dead time increased with the injected activity. A dead-time loss greater than 5% affected 14.4% and 5.7% of QSPECT scans performed at days 1 and 3, respectively. This resulted in renal absorbed dose estimates that would have been underestimated by more than 5% in 5.7% of cycles if no dead-time correction was applied, with a maximum underestimation of 22.1%. In the simulated empiric regime, this potential dose underestimation would have been limited to 6.2%.
Conclusion:
Dead-time correction improves the accuracy of dosimetry in 177Lu radionuclide therapy and is warranted in personalized PRRT.
More Related Videos
09:44Pretargeted Radioimmunotherapy Based on the Inverse Electron Demand Diels-Alder Reaction
Published on: January 29, 2019
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Noncompartmental Analysis: Mean Residence Time
After the administration of a drug through intravenous bolus injection, the drug molecules are distributed throughout the body and remain there for varying periods. The MRT represents the average time these drug molecules stay in the...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET