Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
A demonstration of high field-of-view stability in hands-free echocardiography
Marloes Sjoerdsma1, Louis S Fixsen2, Thijs Schoots2,3
1Department of Biomedical Engineering, Eindhoven University of Technology, Groene Loper, Building 15, Eindhoven, The Netherlands. m.sjoerdsma@tue.nl.
Background:
Exercise stress echocardiography is clinically used to assess cardiovascular diseases. For accurate cardiac evaluation, a stable field-of-view is required. However, transducer orientation and position are difficult to preserve. Hands-free acquisitions might provide more consistent and reproducible results. In this study, the field-of-view stability and variability of hands-free acquisitions are objectively quantified in a comparison with manually obtained images, based on image structural and feature similarities. In addition, the feasibility and consistency of hands-free strain imaging is assessed.
Methods:
In twelve healthy males, apical and parasternal images were acquired hands-free, using a fixation device, and manually, during semi-supine exercise sessions. In the final ten seconds of every exercise period, the image structural similarity and cardiac feature consistency were computed using a steerable pyramid employing complex, oriented wavelets. An algorithm discarding images displaying lung artifacts was created. Hands-free strain consistency was analyzed.
Results:
Hands-free acquisitions were possible in 9 of the 12 subjects, whereas manually 10 out of 12 could be imaged. The image structural similarity was significantly improved in the hands-free apical window acquisitions (0.91 versus 0.82), and at least equally good in the parasternal window (0.90 versus 0.82). The change in curvature and orientation of the interventricular septum also appeared to be lower in the hands-free acquisitions. The variability in field-of-view was similar in both acquisitions. Longitudinal, septal strain was shown to be at least as consistent when obtained hands-free compared to manual acquisitions.
Conclusions:
The field-of-view was shown to be more or equally stable and consistent in the hands-free data in comparison to manually obtained images. The variability was similar, thus respiration- and exercise-induced motions were comparable for manual and hands-free acquisitions. Additionally, the feasibility of hands-free strain has been demonstrated. Furthermore, the results suggest the hands-free measurements to be more reproducible, though further analysis is required.
More Related Videos
09:05Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...