The diagnostic accuracy of left ventricular ejection fraction assessment between visual and artificial
Nikola Kolobaric1,2, Nickolas Beauregard1, William Barbour1,2
1Division of Cardiology, CAPITAL Research Group, University of Ottawa Heart Institute, Ottawa, ON, Canada.
Background:
Focused cardiac ultrasound (FoCUS) has become the standard of care for bedside assessments of cardiac function. With the integration of artificial intelligence (AI), there is limited evidence comparing it to bedside visual assessments by experienced users.
Methods:
In our prospective study conducted at Tufts Medical Center in Boston, Massachusetts from December 2020 to March 2022, patients ≥ 18 years requiring a TTE were recruited by convenience sampling. They each underwent FoCUS LVEF classification by AI and bedside visual assessment, with TTE as reference. LVEF was calculated by Simpson's biplane method of disks in AI-FoCUS and TTE, and visual global assessment by the bedside sonographer. Data analysis was completed in October 2025.
Results:
Our 215 participants had a median age of 63 (IQR 49-73) years with 83 (38.6%) being female. AI-FoCUS assessments showed good agreement with TTE (intraclass correlation coefficient 0.84, 95% CI: 0.80-0.88), while visual assessments had stronger concordance (intraclass correlation coefficient 0.97, 95% CI: 0.96-0.98). Categorization of LV dysfunction severity showed excellent agreement with TTE for both AI-FoCUS (kappa 0.89, 95% CI: 0.84-0.94) and visual estimate (kappa 0.96, 95% CI: 0.92-0.99). For AI-FoCUS, the area under the curve (AUC) for identifying an abnormal LVEF (< 50%) was 0.9779 (95% CI: 0.9591-0.9967) with sensitivity 90.9% (95% CI: 88.1-100), specificity 95.6% (95% CI: 92.6-98.6), positive predictive value (PPV) 0.79 (95% CI: 0.66-0.92) and negative predictive value (NPV) 0.98 (95% CI: 0.96-1.00). For visual estimate, the AUC was 0.9961 (95% CI: 0.9915-1.000) with sensitivity 90.9% (95% CI: 88.1-100), specificity 97.8% (95% CI: 95.7-99.9), PPV 0.88 (95% CI: 0.77-0.99) and NPV 0.98 (95% CI: 0.96-1.00).
Conclusion:
AI-assisted FoCUS LVEF assessments provide accurate estimates for the presence and severity of LV dysfunction but are outperformed in the latter by experienced bedside users.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...