Radiation protection of operators and patients in a hybrid Angio-MR suite
M Dehairs1, N W Marshall1, H Bosmans2
1UZ Gasthuisberg, Department of Radiology, Herestraat 49, 3000 Leuven, Belgium.
Abstract:
This work investigates the patient eye lens dose and x-ray scatter to the operator expected for a proposed hybrid Angio-MR concept. Two geometries were simulated for comparative assessment: a standard C-arm device for neuro-angiography applications and an innovative hybrid Angio-MR system concept, proposed by Siemens Healthineers. The latter concept is based on an over-couch x-ray tube and a detector inside an MRI system, with the aim of allowing combined, simultaneous MRI and x-ray imaging for procedures such as neurovascular interventions (including x-ray fluoroscopy and angiography imaging, 3D imaging, diffusion, and perfusion). To calculate the scattered radiation dose to the physician, Monte Carlo simulations were performed. Dose estimates of simplified models of the brain and eyes of both the patient and the physician and of the physician's torso and legs have been calculated. A number of parameters were varied in the simulation including x-ray spectrum, field of view (FOV), x-ray tube angulation, presence of shielding material and position of the physician. Additionally, 3D dose distributions were calculated in the vertical and horizontal planes in both setups. The patient eye lens dose was also calculated using a detailed voxel phantom and measured by means of thermoluminescent dosimeters (TLDs) to obtain a more accurate estimate. Assuming the same number of x-rays and the same size of the irradiated area on the patient's head, the results show a significant decrease in the scattered radiation to the physician for the Angio-MR system, while large increases, depending on setup, are expected to patient eye lens dose.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT