Task-based selection of three-dimensional rotational angiography imaging modes using in-house phantom
L E Lubis1, R A Basith2, I Hariyati3
1Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Kampus UI Depok 16424, Indonesia; Radiology Unit, Universitas Indonesia Hospital, Kampus UI Depok 16424, Indonesia.
Radiography (London, England : 1995)
|April 11, 2024
Summary
The "head limited" mode in three-dimensional rotational angiography (3DRA) is recommended for cranial procedures. It offers a balance of lower radiation dose and sufficient image quality for contrast-enhanced imaging.
Area of Science:
- Medical Imaging
- Radiology
- Diagnostic Imaging
Background:
- Clinicians and radiographers face challenges selecting appropriate three-dimensional rotational angiography (3DRA) modes for cranial applications due to similar protocol names.
- Ambiguity in 'cerebral' and 'head limited' 3DRA modes complicates optimal use.
Purpose of the Study:
- To evaluate and recommend the optimal mode selection for cranial 3DRA procedures.
- To address clinical difficulties in choosing between 'cerebral' and 'head limited' 3DRA modes.
Main Methods:
- An in-house phantom simulating cranial vessels with varying iodinated contrast agent concentrations was utilized.
- Image quality was assessed using signal difference to noise ratio (SDNR), and radiation dose was measured via Kerma-area product (KAP).
- X-ray spectrum analysis provided quantitative evaluation.
Main Results:
- The 'head limited' mode is suggested for clinical use.
- The 'low' detail option within 'head limited' mode yields the lowest KAP and slightly superior SDNR compared to the 'normal' detail option.
- A minimum iodine concentration of 18.5 mg/ml is needed with the 'low' detail option for comparable SDNR.
Conclusions:
- The 'head limited' mode with 'low' detail settings is advisable for contrast-enhanced 3DRA procedures.
- Collaboration between clinicians, physicists, and manufacturers is crucial for standardized 3DRA mode selection.
- Objective dose and image quality metrics reduce subjectivity in 3DRA mode selection.