Performance characteristics of thromboelastometry assays using incompletely filled and prolonged stored samples
Helene M Paarup1, Mads H Rasmussen1, Mark H Yazer1,2
1Department of Clinical Immunology, Odense University Hospital, Odense, Denmark.
Transfusion
|June 2, 2020
Summary
Thromboelastometry (TEM) testing is reliable for fully filled or half-filled samples, and for samples stored at 4°C. However, samples less than half-filled yield abnormal results and should be avoided for accurate TEM analysis.
Area of Science:
- Clinical diagnostics
- Hemostasis and thrombosis
- Transfusion medicine
Background:
- Thromboelastometry (TEM) is crucial for guiding transfusion therapy in patients with severe bleeding.
- Investigating the impact of sample filling volume and storage conditions on TEM accuracy is essential for reliable clinical application.
Purpose of the Study:
- To evaluate the effect of incompletely filled samples on Thromboelastometry (TEM) results.
- To assess the impact of prolonged cold storage (4°C) on TEM parameters.
Main Methods:
- Whole blood samples from healthy donors were diluted to achieve varying blood:citrate ratios (90:10 to 40:60).
- Samples were tested using EXTEM, INTEM, and FIBTEM assays on ROTEM delta machines.
- Additionally, samples stored at 4°C for 16-20 hours were retested.
Main Results:
- Samples with 90:10 and 80:20 dilutions (fully and half-filled) showed ROTEM parameters within reference ranges.
- Samples with 70:30 to 40:60 dilutions (less than half-filled) produced abnormal TEM results.
- TEM results for fully filled samples stored at 4°C for 16-20 hours remained within reference intervals.
Conclusions:
- Thromboelastometry (TEM) provides normal results for completely filled and half-filled samples.
- Prolonged cold storage (4°C) of fully filled samples does not significantly affect TEM results.
- Samples less than half-filled should not be used for Thromboelastometry (TEM) testing due to unreliable results.