Related Experiment Video
Updated: Dec 20, 2025

The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Spontaneous Device Detachment After its Partial Deployment During Left Atrial Appendage Occlusion: A Nightmare in the
Giuseppe Talanas1, Eleonora Moccia, Giuseppe D Sanna
1Clinical and Interventional Cardiology, Sassari University Hospital, Via Enrico De Nicola 14, 07100 Sassari, Italy. giuseppe.talanas@aousassari.it.
Abstract:
To date, this is the first description of a spontaneous unscrewing of an Amplatzer Amulet device from its delivery cable while its lobe was partially deployed within the left atrial appendage. In such cases, the device should be pushed forward to complete the deployment and to rescrew the device to the DC in order to prevent an unavoidable cardiac surgery.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization