Fast Fourier transform combined with phase leading compensator for respiratory motion compensation system
Chia-Chun Kuo1,2,3, Ho-Chiao Chuang4, Ai-Ho Liao5,6
1Department of Radiation Oncology, Taipei Medical University Hospital, Taipei, Taiwan.
Background:
The reduction of the delaying effect in the respiratory motion compensation system (RMCS) is still impossible to completely correct the respiratory waveform of the human body due to each patient has a unique respiratory rate. In order to further improve the effectiveness of radiation therapy, this study evaluates our previously developed RMCS and uses the fast Fourier transform (FFT) algorithm combined with the phase lead compensator (PLC) to further improve the compensation rate (CR) of different respiratory frequencies and patterns of patients.
Methods:
In this study, an algorithm of FFT automatic frequency detection was developed by using LabVIEW software, uisng FFT combined with PLC and RMCS to compensate the system delay time. Respiratory motion compensation experiments were performed using pre-recorded respiratory signals of 25 patients. During the experiment, the respiratory motion simulation system (RMSS) was placed on the RMCS, and the pre-recorded patient breathing signals were sent to the RMCS by using our previously developed ultrasound image tracking algorithm (UITA). The tracking error of the RMCS is obtained by comparing the encoder signals of the RMSS and RMCS. The compensation effect is verified by root mean squared error (RMSE) and system CR.
Results:
The experimental results show that the patient's respiratory patterns compensated by the RMCS after using the proposed FFT combined with PLC control method, the RMSE is between 1.50-5.71 and 3.15-8.31 mm in the right-left (RL) and superior-inferior (SI) directions, respectively. CR is between 72.86-93.25% and 62.3-83.81% in RL and SI, respectively.
Conclusions:
This study used FFT combined with PLC control method to apply to RMCS, and used UITA for respiratory motion compensation. Under the automatic frequency detection, the best dominant frequency of the human respiratory waveform can be determinated. In radiotherapy, it can be used to compensate the tumor movement caused by respiratory motion and reduce the radiation damage and side effects of normal tissues nearby the tumor.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Fast Fourier Transform
The computational efficiency of the FFT becomes...
Discrete Fourier Transform
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Continuous -time Fourier Transform
Properties of Fourier Transform I
In radio broadcasting, multiple audio signals often need to be transmitted simultaneously. The Fourier...
Basic signals of Fourier Transform
The sinc function, defined as sinc(x) = sin(πx)/(πx), is particularly notable for its symmetry and behavior at...