Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
Ventricular strain analysis in patients with no structural heart disease using a vendor-independent speckle-tracking
Hongmei Xia1, Darwin F Yeung2, Cristina Di Stefano3
1Department of Ultrasound, Xinqiao Hospital, Third Military Medical University, Chongqing, China.
Background:
Ventricular strain measurements vary depending on cardiac chamber (left ventricle [LV] or right ventricle [RV]), type of strain (longitudinal, circumferential, or radial), ventricular level (basal, mid, or apical), myocardial layer (endocardial or epicardial), and software used for analysis, among other demographic factors such as age and gender. Here, we present an analysis of ventricular strain taking all of these variables into account in a cohort of patients with no structural heart disease using a vendor-independent speckle-tracking software.
Methods:
LV and RV full-thickness strain parameters were retrospectively measured in 102 patients (mean age 39 ± 15 years; 62% female). Within this cohort, we performed further layer-specific strain analysis in 20 subjects. Data were analyzed for global and segmental systolic strain, systolic strain rate, early diastolic strain rate, and their respective time-to-peak values.
Results:
Mean LV global longitudinal, circumferential, and radial strain values for the entire cohort were - 18.4 ± 2.0%, - 22.1 ± 4.1%, and 43.9 ± 12.1% respectively, while mean RV global and free wall longitudinal strain values were - 24.2 ± 3.9% and - 26.1 ± 5.2% respectively. Women on average demonstrated higher longitudinal and circumferential strain and strain rate than men, and longer corresponding time-to-peak values. Longitudinal strain measurements were highest at the apex compared with the mid ventricle and base, and in the endocardium compared with the epicardium. Longitudinal strain was the most reproducible measure, followed closely by circumferential strain, while radial strain showed suboptimal reproducibility.
Conclusions:
We present an analysis of ventricular strain in patients with no structural heart disease using a vendor-independent speckle-tracking software.