Reproducibility of manual choroidal thickness measurements using optical coherence tomography
M M Esteban-Ortega1, M Steiner2, I García-Lozano1
1Sección de Oftalmología, Hospital Universitario Infanta Sofía, Madrid, España; Departamento de Medicina, Facultad de Ciencias Biomédicas, Universidad Europea, Madrid, España.
Archivos De La Sociedad Espanola De Oftalmologia
|June 14, 2020
Summary
Manual optical coherence tomography (OCT) measurements of choroidal thickness (CT) demonstrate good reproducibility. These findings support OCT
Area of Science:
- Ophthalmology
- Medical Imaging
Background:
- Spectral-domain optical coherence tomography (SD-OCT) is crucial for measuring choroidal thickness (CT).
- Elevated CT is associated with various ocular and systemic diseases.
- Concerns exist regarding the reproducibility and external validity of OCT measurements.
Purpose of the Study:
- To evaluate the inter-observer and intra-observer variability of manual choroidal thickness (CT) measurements using spectral-domain optical coherence tomography (SD-OCT).
- To assess the reliability of manual OCT for multicenter studies.
Main Methods:
- Manual CT measurements were performed on 40 eyes from 21 subjects (11 healthy, 10 with ankylosing spondylitis) using RTVue-100 OCT.
- Nine independent ophthalmologists from six centers conducted the measurements.
- Inter-observer variability was assessed using the intra-class correlation coefficient (ICC); intra-observer variability was assessed in two ophthalmologists.
Main Results:
- The mean subfoveal choroidal thickness (CT) was 364.9±85.1μm.
- The inter-observer ICC was 0.823, indicating good agreement among observers.
- Intra-observer ICCs were high, at 0.885 and 0.925, demonstrating excellent consistency for individual observers.
Conclusions:
- Manual choroidal thickness (CT) measurements using spectral-domain optical coherence tomography (SD-OCT) exhibit good concordance.
- The results suggest that manual OCT measurements are a valid and reliable tool for multicenter research.
- This supports the use of manual OCT in clinical trials and large-scale studies investigating choroidal thickness.