The assessment of esophageal pressure using different devices: a validation study
Davide Chiumello1,2,3, Alessio Caccioppola4, Tommaso Pozzi4
1Department of Anesthesia and Intensive Care, ASST Santi Paolo e Carlo, San Paolo University Hospital, Milan, Italy - davide.chiumello@unimi.it.
New systems for measuring esophageal pressure, including Optivent and bedside equipment, are clinically acceptable for managing acute respiratory distress syndrome (ARDS) patients. This could expand the use of esophageal pressure monitoring in clinical practice.
Area of Science:
- Critical Care Medicine
- Respiratory Physiology
- Medical Device Technology
Background:
- Esophageal pressure measurement aids in ventilatory management for acute respiratory distress syndrome (ARDS) but is primarily used in research.
- Accurate esophageal pressure monitoring is crucial for optimizing mechanical ventilation strategies.
Purpose of the Study:
- To compare measurements of end-expiratory esophageal pressure, transpulmonary pressure, and lung stress using a manual gold standard device, a new automatic system (Optivent), and a bedside system.
- To assess the clinical acceptability and potential for wider application of non-research-grade esophageal pressure measurement tools.
Main Methods:
- Esophageal pressure was measured in 40 sedated, paralyzed, mechanically ventilated ARDS patients at three positive end-expiratory pressure (PEEP) levels.
- Measurements were compared between a dedicated manual device (gold standard), the Optivent system, and a bedside system (ventilator + hemodynamic monitor).
Main Results:
- Mean esophageal pressures measured by the dedicated, bedside, and Optivent systems were 10.0±4.2, 10±4, and 9.9±4.0 cmH2O, respectively.
- Bias and limits of agreement indicated clinical acceptability for both Optivent and the bedside system compared to the gold standard for esophageal pressure, transpulmonary pressure, and lung stress.
- Specific bias values for esophageal pressure were 0.2 cmH2O (Optivent) and -0.1 cmH2O (bedside).
Conclusions:
- Both the Optivent system and the bedside equipment demonstrated clinical acceptability when compared to a gold standard device.
- The availability and validation of these systems may facilitate broader clinical adoption of esophageal pressure monitoring in ARDS management.
- Wider use of esophageal pressure measurement can potentially improve ventilatory strategies and patient outcomes in ARDS.
More Related Videos
06:19Author Spotlight: Implications of Non-Nutritive Sucking on Speech Emergence and Infant Development
Published on: April 19, 2024
08:25Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...