AVERAGE GLANDULAR DOSES AND NATIONAL DIAGNOSTIC REFERENCE LEVELS IN MAMMOGRAPHY EXAMINATIONS IN TURKEY
Aydın Parmaksız1, Gökçe K Ataç2, Emine Bulur1
1Radiation and Accelerator Technologies Department, Turkish Atomic Energy Authority, Atom Street, 06983 Kahramankazan, Ankara, Turkey.
Abstract:
In order to establish national diagnostic reference levels for mammography examinations, entrance skin air kerma, entrance skin dose and average glandular doses (AGDs) were calculated for a total of 25 624 cranio-caudal (CC) and mediolateral oblique (MLO) projections of 6309 patients for 40-49 and 50-64 age groups. The average entrance skin air kerma and entrance skin dose values for both age groups were found to be higher in MLO projections compared with CC projections. The minimum and maximum values of AGDs were determined as 0.4 and 7.9 mGy for both projections. The maximum numbers of AGDs for CC and MLO projections were calculated in the range of 1.1-1.5 and 1.6-2.0 mGy, respectively. The third quartile values of AGDs were calculated for each compressed breast thickness between 20 and 99 mm. The first national diagnostic reference levels of the country were established for each 10-mm compressed breast thickness in mammography examinations.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
13:43Intraductal Delivery and X-ray Visualization of Ethanol-Based Ablative Solution for Prevention and Local Treatment of Breast Cancer in Mouse Models
Published on: April 1, 2022
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT