Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
Fluorous-Phase Ion-Selective pH Electrodes: Electrode Body and Ionophore Optimization for Measurements in the
Xin V Chen1, Maral P S Mousavi1, Philippe Bühlmann1
1Department of Chemistry, University of Minnesota, 207 Pleasant St. SE, Minneapolis, Minnesota 55455, United States.
Abstract:
Because of their low polarity and polarizability, fluorous sensing membranes are both hydrophobic and lipophobic and exhibit very high ion selectivities. Here, we report on a new fluorous-membrane ion-selective electrode (ISE) with a wide sensing range centered around physiologically relevant pH values. The fluorophilic tris[perfluoro(octyl)butyl]amine (N[(CH2)4Rf8]3) was synthesized and tested as a new H+ ionophore using a redesigned electrode body that provides excellent mechanical sealing and much improved measurement reliability. In a challenging 1 M KCl background, these fluorous-phase ISEs exhibit a sensing range from pH 2.2 to 11.2, which is one of the widest working ranges reported to date for ionophore-based H+ ISEs. High selectivities against common interfering ions such as K+, Na+, and Ca2+ were determined (selectivity coefficients: logK H, K pot = - 11.6; logK H, Na pot = - 12.4; logK H, Ca pot < - 10.2). The use of the N[(CH2)4Rf8]3 ionophore with its -(CH2)4- spacers separating the amino group from the strongly electron-withdrawing perfluorooctyl groups improved the potentiometric selectivity as compared to the less basic tris[perfluoro(octyl)propyl]amine ionophore. The use of N[(CH2)4Rf8]3 also made the ISE less prone to counter anion failure (i.e., Donnan failure) at low pH than the use of tris[perfluoro(octyl)pentyl]amine with its longer -(CH2)5- spacers, which more effectively shield the amino center from the perfluorooctyl groups. In addition, we exposed both conventional plasticized PVC-phase pH ISEs and fluorous-phase pH ISEs to 10% serum for 5 days. Results show that the PVC-phase ISEs lost selectivity while their fluorous-phase counterparts did not.
More Related Videos
08:17Fluorescent Nanoparticles for the Measurement of Ion Concentration in Biological Systems
Published on: July 4, 2011
07:38Functional Characterization of Na+/H+ Exchangers of Intracellular Compartments Using Proton-killing Selection to Express Them at the Plasma Membrane
Published on: March 30, 2015
Related Concept Videos
Potentiometry: Membrane Electrodes
Fluorescence and Phosphorescence: Instrumentation