Nominal Size and Taper Analysis of Novel Metallurgy NiTi Files
Melita Islambasic1, Marcelo Santos Coelho1, Mary T Pettiette1
1Department of Endodontics, University of North Carolina at Chapel Hill, Chapel Hill, US.
Objective:
The aim of this investigation was to assess the nominal tip diameter, taper and true pilot tip length of three nickel-titanium (NiTi) rotary files before and after use.
Methods:
Three brands of NiTi files of size 25.08 were evaluated: ProTaper Uni-versal (Dentsply Tulsa Specialties, Tulsa, USA), Channels PT (Insight Endo for Henry Schein, Melville, NY) and ProTaper Gold (Dentsply Tulsa Specialties), with ten files from each brand (total n=30). Scanning electron microscope (SEM) images of the files at 50x magnification were acquired before and after the files were used on endodontic training blocks, and the images were analysed by two independent investigators using ImageJ software. The nominal tip diameter (D0), taper and true pilot tip length (measured as the distance from the advertised diameter size of the file to the tip of the file) were recorded for each file and were analysed for statistical differences using repeated measures of analysis of variance (r-ANOVA) and Tukey's post-hoc test (P<0.05).
Results:
The marginal means of the nominal size showed statistically significant differences between brands of the same size; statistically significant differences were also observed between the advertised sizes and the actual sizes (P<0.001). The mean taper values showed statistically significant differences from the advertised (p<0.001) sizes, except for the ProTaper Gold (P=0.023) group. The changes in the true pilot tip length before and after use were not statistically significant (P>0.05). The pilot tip lengths varied between brands.
Conclusion:
All nominal tip diameters and most taper sizes of the analysed brands show statistically significant differences from the nominal sizes advertised by the manufacturers. The differences in true pilot tip lengths between brands indicate a current lack of standardization.
More Related Videos
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
07:47Reverse Microemulsion-mediated Synthesis of Monometallic and Bimetallic Early Transition Metal Carbide and Nitride Nanoparticles
Published on: November 27, 2015
Related Concept Videos
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...