Adaptive transverse blood velocity estimation in medical ultrasound: A simulation study
Fatemeh Makouei1, Babak Mohammadzadeh Asl1
1Department of Biomedical Engineering, Tarbiat Modares University, Tehran, Iran.
Abstract:
Undoubtedly, highly valuable information about vascular anomalies is attained by the examination of the blood flow profile. The chief drawback of the conventional medical ultrasound in preparation of the blood periodogram is the measurement system shortcoming at the beam to flow angles near 90°. Recently, a method based on transverse oscillation (TO) approach, known as "Fourth-order estimation", has been developed to directly estimate the transverse power spectral density (PSD) of the fully transverse blood flow. One of the basic requirements to accomplish acceptable PSDs by this technique is the sufficiently large observation window. In this paper, two adaptive approaches for efficient estimation of the velocity spectrum of a fully transverse flow by a limited observation window length are described. The first proposed adaptive approach is based on the minimum variance adaptive spectral estimation in combination with the well-known TO technique (TO-MV). Then, by exploiting the eigenspace separation of the observed data to eliminate the contribution of the undesired components, the second technique (TO-EIBMV) is developed. The approaches are validated using Field II simulations for pulsating flow. The proposed methods are tested and compared to the conventional TO transverse spectral estimator by metrics of relative standard deviation (RSD) and relative bias (RB). One of the main achievements is the decrement of the required data samples for spectrogram estimation, which leads to a better temporal resolution. Moreover, for the analyzed adaptive techniques, the robustness of the estimation results for the beam to flow angles of 60-90° and vessel depths ranging from 20 mm to 60 mm are investigated.