Towards safe operation of an active retinal prosthesis during functional MRI and diffusion tensor imaging
Ken E Sakaie1, Barry Kuban2, Aaron Fleischman2
1Department of Diagnostic Radiology, Imaging Institute, The Cleveland Clinic, United States of America.
Magnetic Resonance Imaging
|June 28, 2020
Summary
The Argus II retinal prosthesis can safely operate during functional MRI (fMRI) and diffusion tensor imaging (DTI) scans. Induced currents remained below established safety thresholds, allowing for combined use in patients.
Area of Science:
- Medical device engineering
- Neuroscience
- Radiology
Background:
- The Argus II retinal prosthesis restores some vision to patients with retinitis pigmentosa.
- Assessing the safety of medical implants during advanced neuroimaging is crucial.
Purpose of the Study:
- To evaluate the operational capability of the Argus II retinal prosthesis during fMRI and DTI.
- To measure induced currents in the prosthesis to ensure patient safety during these imaging modalities.
Main Methods:
- A modified Argus II retinal prosthesis was used for current measurements.
- The prosthesis was tested within an ASTM phantom during diagnostic, fMRI, and DTI scans.
- Active electronics were adapted for in-scan operation.
Main Results:
- The Argus II maintained communication during fMRI and DTI scans, with minor interruptions during pre-scans.
- Induced current levels were consistently lower during fMRI and DTI compared to diagnostic scans.
- Measurements confirmed induced currents remained within safe limits.
Conclusions:
- Functional MRI and diffusion tensor imaging can be safely conducted with an operating Argus II retinal prosthesis.
- The study supports the combined use of Argus II and advanced neuroimaging techniques.
- This finding expands imaging possibilities for patients with retinal implants.