Concentrated stress effects of contoured and non-contoured high Tibial osteotomy plates: A finite-element study
Pei-Wei Weng1, Chen-Kun Liaw2, Chia-Hsien Chen1
1Department of Orthopaedics, School of Medicine, College of Medicine, Taipei Medical University, Taipei, Taiwan; Department of Orthopaedics, Shuang Ho Hospital, Taipei Medical University, New Taipei City, Taiwan; Graduate Institute of Biomedical Materials and Tissue Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei, Taiwan.
Background:
The TomoFix plate has been extensively used in high tibial osteotomy surgery to stabilize the distracted tibial bones. However, distal pain related to plate irritation was considered one of the most relevant complications for this fixation device. This study aimed to correlate reports of distal pain with the profiles of the distracted tibia and initial plate and plate contour.
Methods:
This study used the finite-element method to investigate the profile-, distraction-, and contour-induced effects on stress distribution of the distal tibia-plate contact. The associations of two tibia profiles (normal and concave), distraction angles, and two plate profiles (contoured and non-contoured) were discussed in this study. The areas and stresses of the distal tibia-plate contact were chosen as comparison indices.
Findings:
Before weight-bearing, the non-contoured plates of the normal and concave tibia profiles consistently showed less contact area at the distal tibia-plate region. Consequently, the physiological loads make the non-contoured plate subject to more concentrated bone stresses and thus may induce more pain at the distal tibia-plate region than a contoured plate. When the distraction angle decreases, the tibia-plate gap increases. Prior to fixation, the tibia-plate gap can be evaluated by the profiles of the distracted tibia and non-contoured plate by use of anteroposterior radiograph and computer-aided simulation.
Interpretation:
In the situations of a lower distraction angle or a large tibia-plate gap, the use of a plate bender or a lag screw is recommended in order to contour the plate for reducing the concentrated stress at the distal tibia-plate region.
More Related Videos
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
05:55Author Spotlight: Developing a Rat Model for Weight-Bearing Intervention to Investigate Osteonecrosis of the Femoral Head
Published on: September 27, 2024
Related Concept Videos
Stress Concentrations
Stress Concentrations
The stress...
Stress Concentrations in Circular Shafts
Transformation of Plane Stress
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....