Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Catheter strategy to ease the procedure and reduce radiation exposure when requiring neck access
Rouven Kubicki1, Johanna Hummel2, René Höhn2
1Department of Congenital Heart Disease and Paediatric Cardiology, University Heart Center Freiburg - Bad Krozingen, Medical Centre - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, Freiburg, Germany Rouven.kubicki@universitaets-herzzentrum.de.
Objectives:
To assess the potential occupational radiation reduction and technical feasibility in patients rotated 180° (upside-down) when requiring neck access for transcervical or trans-subclavian catheterisation.
Methods:
Upside-down positioning is defined as rotating patients in supine position by 180°, so that the feet come to rest where the head would otherwise be. We retrospectively evaluated all these procedures performed between March 2016 and May 2019. Furthermore, two different phantoms (paediatric and adult) were used prospectively to quantify the occupational dose between conventional or upside-down positioning. In this context, ambient dose equivalents were measured using real-time dosimeters. Three different projection angles were applied.
Results:
44 patients with median age and body weight of 1.0 year (range 0-56) and 9.5 kg (range 1.3-74.3) underwent 63 procedures positioned upside-down. This position proved advantageous for practical reasons, since the length of the examination table could be optimally used. Additionally, it resulted in a significantly lower overall ambient dose equivalent for the primary operator (PO) of 94.8% (mean: 2569±807 vs 135±23 nSv; p<0.01) in the adult, and of 65.5% (mean: 351±104 vs 121±56 nSv; p<0.01) in the paediatric phantom, respectively.
Conclusion:
Upside-down positioning facilitates handling in a straightforward manner when access from the neck is required. Moreover, it significantly reduces local radiation exposure for the PO in the paediatric and, most impressively, in the adult phantom.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
05:31Transradial Access Chemoembolization for Hepatocellular Carcinoma Patients
Published on: September 20, 2020
Related Concept Videos
Cardiac Catheterization IV: Nursing Management
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Suctioning the Nasopharyngeal Airway
Equipment Required
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: