Related Experiment Video
Updated: Dec 16, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Fast spot-scanning proton dose calculation method with uncertainty quantification using a three-dimensional
Yusuke Nomura1, Jeff Wang2,3, Hiroki Shirato2,4
1Department of Radiation Medical Science and Engineering, Faculty of Medicine, Hokkaido University, Sapporo 060-8638, Japan.
A new deep learning method using a 3D convolutional neural network (3D-CNN) accurately calculates proton therapy doses in near real-time. This approach also estimates dose uncertainties, improving efficiency and applicability for image-guided proton therapy.
Area of Science:
- Medical Physics
- Radiotherapy
- Artificial Intelligence in Medicine
Background:
- Proton therapy offers precise dose delivery but requires accurate dose calculation.
- Real-time dose calculation and uncertainty estimation are crucial for advanced applications like image-guided proton therapy.
- Current methods can be computationally intensive, limiting near real-time applications.
Purpose of the Study:
- To develop a near-real-time spot-scanning proton dose calculation method using a 3D convolutional neural network (3D-CNN).
- To incorporate probabilistic uncertainty estimation into the dose calculation process.
- To evaluate the accuracy, efficiency, and adaptability of the 3D-CNN model for proton therapy dose prediction.
Main Methods:
- A 3D-CNN model was trained and tested on CT images and clinical target volumes from 215 head and neck cancer patients.
- Spot beam data was converted into a 'peak map' (PM) input for the 3D-CNN.
- Transfer learning was employed to adapt the model to different beam parameters and calculation algorithms with minimal data.
Main Results:
- The 3D-CNN model achieved accurate 3D proton dose distribution calculations with a mean absolute error of 0.778 cGyE.
- Predicted uncertainties correlated with dose errors at high contrast edges.
- Inference time was approximately 0.8 seconds per plan on a consumer-grade GPU, demonstrating near real-time capability.
Conclusions:
- The developed 3D-CNN method provides accurate and efficient near-real-time proton dose calculation with uncertainty estimation.
- The model is adaptable to various beam configurations and calculation settings via transfer learning.
- This approach holds significant potential for dose verification, image-guided proton therapy, and other clinical applications.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022