Study on the bending behavior of biodegradable metal cerebral vascular stents using finite element analysis
Weiliang Shi1, Hongxia Li1, Tingzhun Zhu2
1Key Laboratory for Precision & Non-traditional Machining Technology of Ministry of Education, Dalian University of Technology, Dalian, Liaoning 116023, China.
Abstract:
Excellent bending behavior is evaluated as the primary factor during the design of biodegradable metal cerebral vascular stents (BMCVSs), which enables vascular stents to be successfully delivered to the targeted location and avoids unnecessary damage to blood vessels. Unfortunately, this bending behavior has been barely investigated which limits the design of BMCVSs with optimal structures. Herein, six BMCVSs were designed and their bending process were simulated using finite element analysis (FEA). Then, the effects of the stent bridge connection type and structure on the bending behavior were systematically analyzed and an universal mathematical model was further established, in which the influence of the structure parameters of the stent bridge on the flexibility of stents was considered. After that, the bending mechanism of the high-stress zone of the bridge was investigated. Finally, the causes and effects of the self-contacting phenomenon as well as the inner-stent protrusion phenomenon in the bending state were analyzed theoretically, and corresponding solutions were proposed to optimize the design of stents. The numerical results show that the stents with the dislocation-line W-shaped unit have better flexibility than the other stents. The flexibility is positively correlated to the cube of the length of linear part and to the square of the curvature of curved part. The self-contacting phenomenon of the bridge during bending can constrain the formation of inner-stent protrusion, which can eliminate the negative effects of the implanted stents on the hemodynamics in blood vessels. This study is expected to provide practical guidance for the structural design of BMCVSs for clinical applications.
Related Concept Videos
Bending of Material: Problem Solving
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Members Made of Elastoplastic Material
As the bending moment...
Bending of Curved Members - Strain Analysis
The important part of bending analysis for such a member...
Residual Stresses in Bending