Validation of T2 -based oxygen extraction fraction measurement with 15 O positron emission tomography
Dengrong Jiang1,2, Shengwen Deng3, Crystal G Franklin3
1The Russell H. Morgan Department of Radiology & Radiological Science, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Purpose:
To evaluate the accuracy of T2 -based whole-brain oxygen extraction fraction (OEF) estimation by comparing it with gold standard 15 O-PET measurements.
Methods:
Sixteen healthy adult subjects underwent MRI and 15 O-PET OEF measurements on the same day. On MRI, whole-brain OEF was quantified by T2 -relaxation-under-spin-tagging (TRUST) MRI, based on subject-specific hematocrit. The TRUST OEF was compared to the whole-brain averaged OEF produced by 15 O-PET. Agreement between TRUST and 15 O-PET whole-brain OEF measurements was examined in terms of intraclass correlation coefficient (ICC) and in absolute OEF values. In a subset of 10 subjects, test-retest reproducibility of whole-brain OEF was also evaluated and compared between the two modalities.
Results:
Across the 16 subjects, the mean whole-brain OEF of TRUST and 15 O-PET were 36.44 ± 4.07% and 36.45 ± 3.65%, respectively, showing no difference between the two modalities (P = .99). TRUST whole-brain OEF strongly correlated with that of 15 O-PET (N = 16, ICC = 0.90, P = 4 × 10-7 ). The coefficient-of-variation of TRUST and 15 O-PET whole-brain OEF measurements were 1.79 ± 0.67% and 2.06 ± 1.55%, respectively, showing no difference between the two modalities (N = 10, P = .64). Further analyses on the effect of hematocrit revealed that correlation between PET OEF and TRUST OEF with assumed hematocrit remained significant (ICC = 0.8, P < 2 × 10-5 ).
Conclusion:
Whole-brain OEF measured by TRUST was in excellent agreement with gold standard 15 O-PET, with highly comparable accuracy and reproducibility. These findings suggest that TRUST MRI can provide accurate quantification of whole-brain OEF noninvasively.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...