Related Experiment Video
Updated: Dec 15, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Reducing bias in DREAM flip angle mapping in human brain at 7T by multiple preparation flip angles
Hampus Olsson1, Mads Andersen2, Gunther Helms1
1Dept. of Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, SE, Sweden.
Abstract:
DREAM (Dual Refocusing Echo Acquisition Mode) is an ultra-fast multi-slice B1+-mapping technique based on the single-shot STEAM sequence. To study systematic errors at high actual flip angles (FA) and low SNR, DREAM B1+ maps at 3.75×3.75×3.50 mm3 resolution were acquired at 7T in phantoms and human brain in vivo with nominal FAs between 20° and 90° for the two STEAM preparation pulses. Predicted B1+ estimates were underestimated at actual FAs above 50° while noise was prominent below 20°. With a reliable interval of the actual FA between 20° and 50° identified, a B1+ range of 33% - 200% of nominal FA is covered by varying the nominal preparation angle through 25°, 40°, and 60°. Individual B1+ maps are thresholded according to the identified interval and combined into a single map. We demonstrate the benefit of the combined low-noise, low-bias B1+ maps for dual flip angle T1-mapping.

