Factors Influencing the Long-Term Stability of Electronic Tongue and Application of Improved Drift Correction Methods
Zoltan Kovacs1, Dániel Szöllősi2, John-Lewis Zinia Zaukuu1
1Department of Physics and Control, Faculty of Food Science, Szent István University, Somlói út 14-16, H-1118 Budapest, Hungary.
Abstract:
Temperature, memory effect, and cross-contamination are suspected to contribute to drift in electronic tongue (e-tongue) sensors, therefore drift corrections are required. This paper aimed to assess the disturbing effects on the sensor signals during measurement with an Alpha Astree e-tongue and to develop drift correction techniques. Apple juice samples were measured at different temperatures. pH change of apple juice samples was measured to assess cross-contamination. Different sequential orders of model solutions and apple juice samples were applied to evaluate the memory effect. Model solutions corresponding to basic tastes and commercial apple juice samples were measured for six consecutive weeks to model drift of the sensor signals. Result showed that temperature, cross-contamination, and memory effect influenced the sensor signals. Three drift correction methods: additive drift correction based on all samples, additive drift correction based on reference samples, and multi sensor linear correction, were developed and compared to the component correction in literature through linear discriminant analysis (LDA). LDA analysis showed all the four methods were effective in reducing sensor drift in long-term measurements but the additive correction relative to the whole sample set gave the best results. The results could be explored for long-term measurements with the e-tongue.
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Voltammetry: Factors Affecting Measurements
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...