Comparison between different approaches to evaluate fibrous cap thickness in sequential optical coherence tomography
Enrico Romagnoli1,2, Giulia Paoletti1,3, Valeria Marco1
1C.L.I. Foundation, Rome, Italy.
Minerva Cardiology and Angiology
|July 14, 2020
Summary
Optical coherence tomography (OCT) provides reproducible measurements of coronary artery plaque components. Longitudinal assessment of fibrous cap thickness (FCt) using OCT is highly reliable for serial evaluations.
Area of Science:
- Cardiovascular Imaging
- Medical Device Technology
- Interventional Cardiology
Background:
- Optical coherence tomography (OCT) is an intravascular imaging modality.
- Assessing lumen area (LA) and plaque components like lipid arc and fibrous cap thickness (FCt) is crucial for cardiovascular disease management.
- Reproducibility of OCT measurements is essential for reliable serial assessments.
Purpose of the Study:
- To evaluate the reproducibility of lumen area (LA) and plaque component measurements using OCT.
- To compare different methods for assessing fibrous cap thickness (FCt) reproducibility.
- To determine the reliability of OCT for quantitative plaque analysis in coronary arteries.
Main Methods:
- In-vivo human study involving two repeated OCT pullbacks from the same diseased coronary segments.
- Assessment of lumen area (LA), lipid arc, and fibrous cap thickness (FCt).
- Comparison of four distinct approaches for minimal FCt measurement, including longitudinal and segmental analyses.
Main Results:
- Suboptimal in-segment correlation for LA (ICC 0.731) but good correlation for lipid arc (ICC 0.963).
- High reproducibility for FCt measurement with continuous (longitudinal) assessment.
- Best reproducibility for FCt observed with Longitudinal-spot minimal FCt value and Longitudinal-plot mean FCt (ICC 0.893 and 0.952).
Conclusions:
- OCT demonstrates good reproducibility for quantitative plaque measurements, confirming its reliability for serial assessments.
- Longitudinal measurement in adjacent frames is the most accurate and reproducible method for sequential FCt assessment.
- OCT is a reliable tool for monitoring changes in coronary plaque components over time.