Effect of miscentering and low-dose protocols on contrast resolution in computed tomography head examination
Said Mohd Shaffiq Said Rahmat1, Muhammad Khalis Abdul Karim2, Iza Nurzawani Che Isa3
1Department of Radiology, National Cancer Institute, 4, Jalan P7, Presint 7, 62250, Putrajaya, Wilayah Persekutuan Putrajaya, Malaysia.
Background:
Unoptimized protocols, including a miscentered position, might affect the outcome of diagnostic in CT examinations. In this study, we investigate the effects of miscentering position during CT head examination on the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR).
Method:
We simulate the CT head examination using a water phantom with a standard protocol (120 kVp/180 mAs) and a low dose protocol (100 kVp/142 mAs). The table height was adjusted to simulate miscentering by 5 cm from the isocenter, where the height was miscentered superiorly (MCS) at 109, 114, 119, and 124 cm, and miscentered inferiorly (MCI) at 99, 94, 89, and 84 cm. Seven circular regions of interest were used, with one drawn at the center, four at the peripheral area of the phantom, and two at the background area of the image.
Results:
For the standard protocol, the mean CNR decreased uniformly as table height increased and significantly differed (p < 0.05) at +20 cm for MCS (435.70 ± 9.39) and -20 cm for MCI (438.91 ± 10.94) from the isocenter. Similarly, significant reductions (p < 0.05) were also noted for SNR for MCS (at +20 cm) and MCI (at -20 cm). For the low dose protocol, both CNR and SNR were significantly reduced (p < 0.05) at table heights of +20 and -20 cm from the isocenter.
Conclusion:
Miscentering is proven to significantly affect the image quality in both low and standard dose protocols for head CT procedure. This study implies that accurate patient centering is one of the approaches that can improve CT optimization practice.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT