3D-printed bolus improves dose distribution for veterinary patients treated with photon beam radiation therapy
Tiffany Wormhoudt Martin1, Mary-Keara Boss1, Susan M LaRue1
1Colorado State University Veterinary Teaching Hospital, Flint Animal Cancer Center, Department of Environmental & Radiological Health Sciences, 300 W. Drake Rd., Fort Collins, Colorado 80523, USA.
Abstract:
Commercial bolus is frequently used to increase dose at the patient's surface for superficial radiotherapy; however, uneven surfaces can create air gaps and discrepancies between prescribed and delivered dose. The purpose of this study was to determine if a customizable, 3D-printed bolus would improve dosimetry compared with a commercial bolus. For each patient, a planned bolus was generated within planning software, then created with 3D-printing. The treatment plan was recalculated with each bolus in situ. When evaluating tumor volumes at prescription, the 3D-printed bolus was closer to prescription compared to the commercial bolus. There was a significant difference in air gaps in patients receiving radiotherapy to the head (P < 0.001) but the difference was not significant for air gaps in caudal body sites (P = 0.05). Overall, the 3D-printed bolus resulted in reduced air gaps, dosimetry closer to prescription, and should be considered for superficial treatment areas of high irregularity.
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...