Quantifying continuous nebulization via high flow nasal cannula and large volume nebulizer in a pediatric model
1Department of Respiratory Care, Children's Health-Children's Medical Center, Dallas, Texas.
Background:
Use of high flow nasal cannula (HFNC) to deliver aerosolized medications to children has gained considerable interest. However, data on continuous albuterol delivery (CAD) via HFNC are lacking. This study quantified CAD via HFNC/vibrating mesh nebulizer (VMN) and large-volume jet nebulizer (LVN) with face mask (FM) in a pediatric model. Aerosol delivery with two HFNC cannula designs were also compared.
Methods:
A pediatric manikin was connected to a lung simulator (Vt = 150 mL, RR = 28 breaths/minute, I:E 1:2.4) via collecting filter at the carina. XL Pediatric and SML Adult HFNC designs were tested to determine optimal cannula design for CAD. VMN was placed Before humidifier (37°C), albuterol (5 mg/mL) was nebulized at 3, 6, and 12 L/minute (n = 3). To compare HFNC/VMN with LVN and FM, albuterol (15 mg/hour) was aerosolized for 3 hours/device (n = 3). LVN was connected to FM and filled with 9 mL of albuterol (5 mg/mL) and 66 mL of normal saline to deliver 25 mL/hour at 13 L/minute. VMN was connected to the infusion pump to deliver 7.5 mL/hr of albuterol (2 mg/mL). Drug eluted from filters was assayed with UV spectrophotometry (276 nm).
Results:
Optimal aerosol delivery occurred at 3 L/minute (12.6% ± 0.5%) with SML Adult HFNC (P = .04). When used for CAD, inhaled drug delivery with HFNC/VMN (2.2 mg/hr ± 0.1, 14.8% ± 0.7%) was significantly greater than LVN and FM (0.48 ± 0.09 mg/hour, 3.2% ± 0.6%) (P = .001).
Conclusions:
Administration of CAD via HFNC/VMN led to a greater than fourfold increase in drug delivery compared to LVN with FM. Optimal aerosol delivery occurred at 3 L/minute with SML Adult HFNC.
More Related Videos
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Drug Dosing: Infants and Children
Inhaled Medications
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...