Evaluation of serum electrolytes measurement through the 6-year trueness verification program in China
Ying Yan1, Yungang Pu1, Jie Zeng1
1National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Beijing Hospital, National, Center of Gerontology, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing, P.R. China.
Objectives:
The National Center for Clinical Laboratories (NCCL) in China initiated a serum electrolyte trueness verification (ETV) program in 2014 for measurement standardization.
Methods:
Every year, two levels of fresh frozen commutable serum samples determined by inductively coupled plasma mass spectrometry (ICP-MS) reference methods were transported to participating clinical laboratories for the measurement of sodium, potassium, calcium and magnesium. Both samples were measured 15 times in 3 days, and the mean values and coefficient variations (CVs) were calculated from the results. The tolerance limits of trueness (bias), precision (CV) and accuracy (TE) based on the biological variation database were used as the evaluation criteria. The overall trend of the ETV program over 6 years was surveyed by calculating the pass rates of the participating laboratories. The mean bias, inter-laboratory CV, and TE of all laboratory results were analysed. Furthermore, homogeneous and heterogeneous systems were compared, and the bias and CV results of mainstream analysis systems were analysed.
Results:
Pass rates of the three quality specifications increased, and the overall mean bias and inter-laboratory CVs decreased. The homogeneous system was superior to the heterogeneous system for calcium and magnesium measurements. For sodium, potassium, calcium and magnesium, the minimum bias corresponded to Hitachi, Siemens, Beckman AU and Roche, respectively. For inter-laboratory robust CVs, no obvious differences were observed between each peer group.
Conclusions:
The commutable ETV materials assigned via reference methods can evaluate the accuracy and reproducibility of an individual laboratory and the calibration traceability and uniformity between laboratories for measurements.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...