Boosting electrochemical immunosensing performance by employing acetaminophen as a peroxidase substrate
Kai Yan1, Al-Monsur Jiaul Haque1, Ponnusamy Nandhakumar1
1Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan, 46241, Republic of Korea.
Abstract:
In horseradish peroxidase (HRP)-based electrochemical immunosensing, an appropriate HRP substrate needs to be chosen to obtain a high electrochemical signal-to-background ratio. This is limited by the unwanted electrochemical reduction of H2O2, oxidation of the substrate, and the slow electrochemical reduction of the product. Herein, we report acetaminophen (AMP) as a new HRP substrate that allows for highly sensitive immunosensing. Electrochemical behavior and immunosensing performance using AMP are compared with those using the most popular HRP substrate, hydroquinone (HQ). To maintain a high electrocatalytic activity even at an electrode modified with an immunosensing layer, an indium tin oxide electrode partially modified with reduced graphene oxide is employed. AMP allows for a higher signal-to-background ratio than HQ, because the formal potential of AMP is 0.28 V higher than that of HQ and the redox reaction of AMP is as reversible as that of HQ, resulting in a lower detection limit in a sandwich-type immunoassay using AMP for thyroid-stimulating hormone detection. The calculated detection limit is ~0.2 pg/mL. The use of AMP as an HRP substrate is especially promising for highly sensitive electrochemical immunoassays.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Amperometry: Overview