Automated oxygen control in preterm infants, how does it work and what to expect: a narrative review
Hylke H Salverda1, Sophie J E Cramer2, Ruben S G M Witlox2
1Neonatology, Leiden University Medical Center, Leiden, Zuid-Holland, The Netherlands H.H.Salverda@lumc.nl.
Background:
Automated oxygen control systems are finding their way into contemporary ventilators for preterm infants, each with its own algorithm, strategy and effect.
Objective:
To provide guidance to clinicians seeking to comprehend automated oxygen control and possibly introduce this technology in their practice.
Method:
A narrative review of the commercially available devices using different algorithms incorporating rule-based, proportional-integral-derivative and adaptive concepts are described and explained. An overview of how they work and, if available, the clinical effect is given.
Results:
All algorithms have shown a beneficial effect on the proportion of time that oxygen saturation is within target range, and a decrease in hyperoxia and severe hypoxia. Automated oxygen control may also reduce the workload for bedside staff. There is concern that such devices could mask clinical deterioration, however this has not been reported to date.
Conclusions:
So far, trials involving different algorithms are heterogenous in design and no head-to-head comparisons have been made, making it difficult to differentiate which algorithm is most effective and what clinicians can expect from algorithms under certain conditions.
Related Concept Videos
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...