Radionuclide calibrator responses for 224Ra in solution and adsorbed on calcium carbonate microparticles
Elisa Napoli1, Jeffrey T Cessna2, Leticia Pibida2
1Physical Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD, 20899-8462, USA; Oncoinvent AS, Oslo, Norway; Institute of Clinical Medicine, University of Oslo, Oslo, Norway; Department of Radiation Biology, Institute for Cancer Research, Oslo University Hospital, Oslo, Norway.
Abstract:
A suspension of 224Ra adsorbed onto CaCO3 microparticles shows promise for α-therapy of intracavitary micro-metastatic diseases. To facilitate accurate activity administrations, geometry-specific calibration factors for commercially available reentrant ionization chambers (ICs) have been developed for 224RaCl2 solutions and 224Ra adsorbed onto CaCO3 microparticles in suspension in ampoules, vials, and syringes. Ampoules and vials give IC responses consistent with each other to <1%. Microparticles attenuation leads to a ≈1% to ≈2.5% reduction in response in the geometries studied.