Improving radiofrequency power and specific absorption rate management with bumped transmit elements in ultra-high
Alireza Sadeghi-Tarakameh1,2,3, Gregor Adriany3, Gregory J Metzger3
1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey.
Purpose:
In this study, we investigate a strategy to reduce the local specific absorption rate (SAR) while keeping constant inside the region of interest (ROI) at the ultra-high field (B0 ≥ 7T) MRI.
Methods:
Locally raising the resonance structure under the discontinuity (i.e., creating a bump) increases the distance between the accumulated charges and the tissue. As a result, it reduces the electric field and local SAR generated by these charges inside the tissue. The at a point that is sufficiently far from the coil, however, is not affected by this modification. In this study, three different resonant elements (i.e., loop coil, snake antenna, and fractionated dipole [FD]) are investigated. For experimental validation, a bumped FD is further investigated at 10.5T. After the validation, the transmit performances of eight-channel arrays of each element are compared through electromagnetic (EM) simulations.
Results:
Introducing a bump reduced the peak 10g-averaged SAR by 21, 26, 23% for the loop and snake antenna at 7T, and FD at 10.5T, respectively. In addition, eight-channel bumped FD array at 10.5T had a 27% lower peak 10g-averaged SAR in a realistic human body simulation (i.e., prostate imaging) compared to an eight-channel FD array.
Conclusion:
In this study, we investigated a simple design strategy based on adding bumps to a resonant element to reduce the local SAR while maintaining inside an ROI. As an example, we modified an FD and performed EM simulations and phantom experiments with a 10.5T scanner. Results show that the peak 10g-averaged SAR can be reduced more than 25%.