Can we use the ankle non-invasive blood pressure during otolaryngologic surgery: an observational study
Sidi Driss El Jaouhari1, Mohamed Meziane1, Jalal Kessouati2
1Department of Anesthesiology and Critical Care, Faculty of Medicine and Pharmacy, Military Hospital Mohamed V, University of Mohamed V Souissi, Rabat, Morocco.
Introduction:
In otolaryngologic surgery, ankle is frequently used for monitoring anesthesia in place of brachial when the patient doesn´t need invasive arterial cannulation. If there is a clinically useful and Predictable link between the two readings in hemodynamic normal patient, this difference during otolaryngologic surgery, was not evaluated. We aimed to investigate the reliability and the acceptability of non invasive blood pressure measurements at the ankle compared to those obtained concurrently at the arm during otolaryngologic surgery.
Methods:
Eighty ASA grade I and II patients who had to be operated under general anesthesia were taken as subjects for our study. Blood pressures were measured simultaneously in the 2 limbs before induction and then every 10 minutes until the end of the surgical procedure. Readings were initiated concurrently. Statistical analysis was performed with PASW Statistics 13.
Results:
There were 41 males (51.2 %) and 39 females (48.8 %). Bland-Altman analysis of mean difference between the ankle and arm (95 % limits of agreement) was -11.47 (- 23.77 to 0.82) mmHg for systolic blood pressure (SBP), -7.89 (-19.16 to 3.36) mmHg for diastolic blood pressure (DBP) and - 9.09 (18.19 to 0.00) mmHg for mean arterial pressure (MAP). Non-parametric analysis showed that 67.5 % of SBP, 46.2 % of DBP and 56.2 % of MAP measurements differed by > 10mmHg.
Conclusion:
Ankle BP cannot be used routinely in otolaryngological surgery. Although, the ankle can be used as an alternative where the arm cannot be used taking into account a difference.
More Related Videos
15:04An Anatomical Study of Nerves at Risk During Minimally Invasive Hallux Valgus Surgery
Published on: February 17, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Aneurysm IV: Nursing Management
Pre-Procedural Guidelines for Assessing Blood Pressure
Peripheral Artery Disease V: Postoperative Nursing Management
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.