Low Detection Rate of Bordetella pertussis Using the BioFire FilmArray Respiratory Panel 2plus
Sophie Guillot1,2, Assaf Mizrahi3,4, Nathalie Armatys1,2
1Biodiversity and Epidemiology of Bacterial Pathogens, Institut Pasteur, Paris, France.
Abstract:
Syndromic respiratory panels are increasingly used worldwide. Their performance for detection of Bordetella pertussis needs to be evaluated. We found that the FilmArray Respiratory Panel 2plus (RP2+) assay, which uses the pertussis toxin promoter target for B. pertussis, can only detect highly charged samples. Negative RP2+ results should not be interpreted as an absence of B. pertussis in clinical samples.
Insights
The FilmArray Respiratory Panel 2plus (RP2+) assay has limited detection of Bordetella pertussis, particularly in non-highly charged samples. Negative results do not rule out pertussis infection.
Area of Science:
- Clinical microbiology
- Infectious disease diagnostics
- Respiratory pathogen detection
Background:
- Syndromic respiratory panels are widely adopted for diagnosing respiratory infections.
- Accurate detection of Bordetella pertussis is crucial for managing whooping cough outbreaks.
- The performance of specific assays, like the FilmArray RP2+, requires thorough evaluation.
Purpose of the Study:
- To evaluate the diagnostic performance of the FilmArray Respiratory Panel 2plus (RP2+) assay for detecting Bordetella pertussis.
- To determine the limitations of the RP2+ assay in identifying B. pertussis in clinical specimens.
Main Methods:
- Evaluation of the FilmArray Respiratory Panel 2plus (RP2+) assay.
- Focus on the detection target for Bordetella pertussis (pertussis toxin promoter).
- Assessment of assay performance across different sample types or charge characteristics.
Main Results:
- The FilmArray RP2+ assay demonstrated limited sensitivity for Bordetella pertussis detection.
- The assay was primarily effective only with highly charged clinical samples.
- Negative RP2+ results may lead to underdiagnosis of B. pertussis.
Conclusions:
- The FilmArray RP2+ assay's ability to detect Bordetella pertussis is constrained.
- Clinicians should exercise caution when interpreting negative RP2+ results for B. pertussis.
- Alternative or confirmatory testing may be necessary for suspected pertussis cases.