Evaluation of four-dimensional cone beam computed tomography ventilation images acquired with two different linear
Noriyuki Kadoya1, Hikaru Nemoto2, Tomohiro Kajikawa3
1Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan.
Abstract:
We evaluated four-dimensional cone beam computed tomography (4D-CBCT) ventilation images (VICBCT) acquired with two different linear accelerator systems at various gantry speeds using a deformable lung phantom. The 4D-CT and 4D-CBCT scans were performed using a computed tomography (CT) scanner, an X-ray volume imaging system (Elekta XVI) mounted in Versa HD, and an On-Board Imager (OBI) system mounted in TrueBeam. Intensity-based deformable image registration (DIR) was performed between peak-exhale and peak-inhale images. VICBCT- and 4D-CT-based ventilation images (VICT) were derived by DIR using two metrics: one based on the Jacobian determinant and one on changes in the Hounsfield unit (HU). Three different DIR regularization values (λ) were used for VICBCT. Correlations between the VICBCT and VICT values were evaluated using voxel-wise Spearman's rank correlation coefficient (r). In case of both metrics, the Jacobian-based VICBCT with a gantry speed of 0.6 deg/sec in Versa HD showed the highest correlation for all the gantry speeds (e.g., λ = 0.05 and r = 0.68). Thus, the r value of the Jacobian-based VICBCT was greater or equal to that of the HU-based VICBCT. In addition, the ventilation accuracy of VICBCT increased at low gantry speeds. Thus, the image quality of VICBCT was affected by the change in gantry speed in both the imaging systems. Additionally, DIR regularization considerably influenced VICBCT in both the imaging systems. Our results have the potential to assist in designing CBCT protocols, incorporating VICBCT imaging into the functional avoidance planning process.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...